Description
* Humidifier is optional
The AirSense 10 AutoSet CPAP Machine with(optional) HumidAir Heated Humidifier builds on the reliable and advanced features of the S9 line of machines by ResMed. It combines the proven technology of the AutoSet algorithm with the integrated comfort of the Optional ClimateLineAir Heated Tube. This is an auto titrating CPAP machine which adjusts the pressure on a breath by breath basis to deliver the minimal pressure needed to maintain the airway.
Weight – Machine with Water Chamber: 2.75 lbs (44 oz.)
Dimensions – Machine with Water Chamber: 10.04 in L x 4.57 in W x 5.91 in H
Weight – Standard Chamber: .54 lb
Weight – Machine with Side Cover (Without Water Chamber): Approx. 2.452 lbs
Dimensions – Machine with Side Cover (Without Water Chamber): Approx. 8 in L x 4.57 in W x 5.91 in H
Dimensions – Travel Bag: 13.24 in L x 10 in W x 5 in H
Weight – Travel Bag Alone: 1.18 lb
Weight – Machine with Humidifier; Power Supply and Cord; SlimLine Tube; Travel Bag: Approximately 5.1 lbs
Weight – Machine with Humidifier; Power Supply and Cord; OPTIONAL ClimateLineAir Tube; Travel Bag: Approximately 5.25 lbs
Sound Level: 26.6 dBA
Water Chamber Capacity: 380 mL
Weight – 90W Power Supply, Cord and Plug: 1.29 lb
Dimensions – 90W Power Supply Only (Not including Cord or Plug): 5 1/2 in x 1 7/8 in x 1 1/4 in
AC Input Range – 90W Power Supply: 100 – 240V, 50 – 60Hz, 1.0 – 1.5A, Class II
DC Output – 90 W Power Supply: 24V — 3.75A
Typical Power Consumption: 53W (57VA)
Peak Power Consumption: 104W (108VA)
Operating Pressure: 4 – 20 cm H2O
Length – SlimLine Tubing: 6 feet
Materials – SlimLine Tubing: Flexible Plastic
Materials – Water Chamber: Injection Molded Plastic, Stainless Steel, Silicone Seal
Materials – Standard Filter: Polyester non-woven fiber
Temperature Range – Optional ClimateLineAir Heated Tube: 60 – 80° F (16 – 30° C)
Operating Temperature: 41° F – 95° F (5° C – 35° C)
Storage and Transport Temperature: -4° F – 140° F (-20° C – 60° C)
Operating Humidity: 10% – 95% Relative Humidity, non-condensing
Storage and Transport Humidity: 5% – 95% relative Humidity, non-condensing
Operating Altitude: Sea Level – 8500 Feet (2,591 M); Air pressure range 1013 hPa – 738 hPa
Maximum Temperature – Heating Plate: 154° F (68° C)
Maximum Temperature – Cut Out: 165° F (74° C)
Maximum Gas Temperature: Less than or Equal to 106° F (? 41° C)
IEC 60601-1:2005 Classification: Class II (Double Insulation), Type BF, Ingress protection IP22
Machine Warranty: 2 Years